Spring Column
DAC Column
Prep system
The Spring Pre Packed Column (SAC)
The Spring Prepacked HPLC column (SAC) does not create void in the column bed even if used for a long time.
Because the resilience of the spring always compresses the column bed.
- The column shall be packed with a column packing device.
However, it is very simple to use a column because it does not require a packing device.
Once installed, the column bed is stable because there is no need to reassemble or disassemble the column.
- A typical prepacked HPLC column may cause void due to long time use.
However, the SAC column does not produce void because the steady resilience of the spring always presses the column bed at a constant pressure. So the performance of the column is long maintained.
- It is economical to charge different sizes of columns with a single spring column packing device.
- This column does not use electricity, so it can be used for explosion-proof areas.
- The piping is much simpler than the DAC column.
- Distribution plate and collecting plate have been developed with independent design and special processing techniques.
As a result, the column performance is excellent.
Column tube |
Inner diameter |
30㎜, 50㎜ |
Length |
650㎜ |
|
The inner wall of processing |
Honing, polishing |
|
The inner wall surface roughness |
Ra ≤ 0.4㎜ |
|
The maximum bed height |
300㎜ |
|
The minimum bed height |
50㎜ |
|
Maximum pressure |
26㎫ |
|
Filter plate |
Aperture |
2㎛ |
Material |
316L power sintered mesh board |
|
Seals |
Material |
PTFE |
Structure |
Spring pre tightening type V sealing ring |
|
Piston |
Material |
316L |
Distribution mode |
Sector |
|
Connection |
Inlet / Outlet |
1/16“ or 1/8” |
* All contact with the liquid material comprising: 316L, PTFE, PEEK