Loading
  • 퀵메뉴 이름 미입력
  • 퀵메뉴 이름 미입력
  • 퀵메뉴 이름 미입력
  • 퀵메뉴 이름 미입력
  • 퀵메뉴 이름 미입력
  • 퀵메뉴 이름 미입력
home Product Prep HPLC Column DAC Column

DAC Column

Spring Column

DAC Column

Prep system




Dynamic Axial Compression Column System (DAC)

The DAC column has been developed for production or research and can be charged automatically.
The DAC column compresses the column bed with a hydraulic cylinder.



Characteristics of DAC Columns

1. Distribution plates and collecting plates have been developed with independent design and special processing techniques. As a result, the column performance is excellent.
2. A typical prep HPLC column may cause void due to long time use. However, the DAC column does not produce void because the piston head compresses the column bed to a constant pressure at all times.
So the performance of the column is long maintained.
3. Using an instrument that automatically supplies slurry, column packing can be done automatically.
4. Power sources can use compressed air and electricity.
Pneumatic pumps can maintain column performance for a long time because it compress the column bed with the pump pressure.
An explosion-proof electric drive motor may be used if the noise is annoying.
5. Packing using a pneumatic pump in a DAC column with a diameter of more than 800mm takes a long time to pack because there are limitations in flow rate and pressure.
However, the hydraulic station can quickly packing the column.
And after packing, use a pneumatic pump to maintain pressure.



Product specification

 Column

tube

 Inner diameter

 50mm

 100mm

 150mm

200mm 

 300mm

400mm 

 600mm

 800mm

 1000mm

 Length (can be

(customized)

 650mm

 800mm

 1200mm

 The maximun bed

hieght

 300mm

 350mm

 600mm

 The minimun bed

height

 50mm

 The inner wall of

processing

  Honing, polishing

 The inner wall

surface rounghness

   Ra ≤ 0.4㎛ 

 Maximum pressure

 10 Mpa

 Filter

plate

 Aperture

 3㎛

 5

 Material

 316L metal sintered mesh plate

 Seals

 Material

 PTFE

 Structure

 Spring loaded V sealing ring

 Piston

 Material

 316 L

 Distribution mode

 Sector

 Power

unit

 Hydraulic station

(pneumatic)

 Air consumption : max. 600 NL/min

 Hydraulic station

(electric)

  AC 3ø 380V, 1.5㎾

 Control

unit

 Pressure relief valve

  Having overpressure protection 

 Switching valve

  Switching cylinder direction (up, down, hold) 

 Pressure regulator

valve

  Adjust the pump output prerssure

Gauge 

  Working air pressure, oil pressure 

 Scram button

  Emergency

 Dimension

 Width (mm)

 600

 730

 790

 800

 880

 1120

 1400

 1900

2650 

 Depth (mm)

 600

 730

 790

 800

 880

 1120

 1400

 1900

2650 

 Height (mm)

 2360

 2360

 2360

 2360

 2900

 2900

 2980

 4280

4800 

 Connection

 Inlet

 1/16"

 1/8"

 1/8"

 1/4"

 3/8"

 1/2"

 3/4"

 1" 

1-1/2" 

 Outlet

 1/16"

 1/8"

 1/8"

 1/4"

 3/8"

 1/2"

 3/4"

 1"

1-1/2" 

 Air

 10㎜ tube quick coupling 








YoungJin Biochrom co.,ltd.

  • Tel +82-31-736-4103
  • Fax +82-70-4275-5345
  • E-mail tax@yjbiochrom.com